62 II
II - 1 Installation Setup
Starting a Non-Linear Error Compensation Table
8 Select ERROR COMPENSATION in Installation Setup, and select
Non-linear by pressing the TYPE soft key.
8 To start a new error compensation table, first press the EDIT TABLE
soft key.
8 All correction points (up to 200) are equally spaced from the start
point. Enter the distance between each of the correction points.
Press the DOWN arrow key.
8 Enter the table’s start point. The start point is measured from the
scale’s reference point. If this distance is not known, you can move
to the location of the start point, and press ENTER, then press
TEACH POSITION.
Pressing ENTER will save the spacing, and start
point information.